| Operating mode | Absorbance, End-Point (Up to 8 standards), Fixed Time, Kinetic, Differential, Turbidimetric (photometric reading) |
|---|---|
| Throughput | Up to 180 tests/hour (mono or bireagent); 300 te with ISE |
| Light source | 12V 20W halogene lamp |
| Spectral Range | 320-690 nm |
| Filter Wavelength | 340, 405, 450, 492, 505, 546, 578, 630 nm, two free position |
| Optics | Solid state detectors,static interferential filters |
| Mode | True Random Access, Walk Away Analyser, Continuous sample loading, Direct reading in the reaction wells at all the wavelength simultaneously |
| Sample number | 75 positions including samples, calibrators, controls, blank |
| Sample container | 1.5-50µl (0.25µl step) |
| Specimen | Serum, Plasma, CSF, Urine, Blood |
| Priority levels | STAT, Pediatric, Normal, Suspended, Rerun |
| Hardware | Dedicated arm, dedicated syringe (500 uL), level sensor on needle, motorized tray |
| Barcode | Optional Manual for all the samples during positioning |
| Automatic dilution | Pre and post dilution performed automatically or on request |
| Worklists | Up to 5 worklists (STAT, pediatric, suspended, rerun, pediatric) |
| Reagent tray | 27 bottles for R1 (50 mL), cooled 9 indipendent bottles (10 mL) for R2, cooled |
| Cooling system | Peltier based convective air forced, integrated for both R1 and R2, stand-by button with indipendent switch |
| Reagent volume | R1 and R2: 20 to 500 µl (0.1 µl steps) |
| R1 hardware | 1000 uL indipendent syringe, dedicated arm, preheating system on needle, liquid detection on needle, cooling system on carousel, motorized tray |
| R2 hardware | 1000 uL indipendent syringe, dedicated arm, preheating system on needle, liquid detection on needle, cooling system on tray, static tray |
| Barcode | Optional, manual for all the reagents during positioning |
| Number of wells | 60 optic bionex plastic reaction cells |
| Reading volume | 180 to 500 uL |
| Well temperature | 37 °C; ± 0.1 °C, controlled by Peltier and microprocessor |
| Reaction Wells | 11 steps washing station |
| Waste | Automatic separation of concentrated and diluted waste |
| R1 needle | Automatic internal and external washing by dedicated wash station |
| R2 needle | Automatic internal and external washing by dedicated wash station |
| Sample needle | Automatic internal and external washing by dedicated wash station |
| Alarm for diluted waste full | By sensor, runtime |
| Alarm for concentrated waste full | By sensor, runtime |
| Alarm for wash solution dry | By sensor, runtime |
| Personal Computer | External. Minimum requirement: CPU 3.0 GHz, 1024 MB RAM, Microsoft Windows XP or 2000 ®, 1 RS-232, 1 LPT, USB |
| Printer | External |
| Communication | USB or RS-232 |
| Power supply | AC 110-230V, autosense, 50/60 Hz |
| Power consumption | AC 110-230V, autosense, 50/60 Hz |
| Working temperature | 15 - 30 °C |
| Patient entry | By workplan or barcode |
| Quality control | Levey-Jennings and Westgaard rules |
| Other features | Run time monitor, continuous sample loading, full spectrum display |
Biochemical Systems International Srl reserves the right to change specifications without prior notice.