GENERAL DESCRIPTION

MULTICARE IN
DIAGNOSTIC AID MULTIPARAMETER WITH CAPILLARY ASPIRATION

TECHNICAL FEATURES

MULTICARE IN
DIAGNOSTIC AID MULTIPARAMETER WITH CAPILLARY ASPIRATION

METER
Measurement range
10-600 mg/dl (0.6-33.3 mmol/l) glucose ELECTRODES
130-400 mg/dl (3,3-10,2 mmol/l) cholesterol
50-500 mg/dl (0,56-5,6 mmol/l) TRIGLYCERIDES
Power Source
2X3V lithium button cell, type CR 2032.
The battery can be used for approx. 1 year with 3 measurements per day (approx. 1000 measurements)
Switch On
Automatic Switch on/off by strip detection
Dimension
Length: 97 mm
Height: 20,5 mm
Width: 49 mm
Weight
65 g (incl. battery)
Indicator
LCD display
Climatic operating condition
Temperature limits: 10 to 40 °C
Humidity limits: 0% to 90% relative humidity (non-condensing)
Altitude limit: 3000 metres
Memory capacity
500 measured values with time and date
Interface for data processing
Connecting cable
Warranty of the instrument
3 years
Contents of the instrument box
1 meter, 1 lancing device, 1 bag of 10 sterile lancets, 1 vial of 10 glucose strips, 1 data-chip, 1 hardcase,1 user's manual, 1 rapid guide, 1 lancing device instructions


GLUCOSE ELECTRODES STRIPS
Sample to be used
Fresh capillary whole blood
Minimum volume: 0.5 µl
Range of Measurement
10-600 mg/dl (0,6-33,3 mmol/l)
Averange time for reading
5 second
Storage conditions
5-30 °C
Table of results
Glucose below 10 mg/dL (0,6 mmol/L) = LO*
Glucose between 10 mg/dL (0,6 mmol/L) and 600 mg/dL (33,3 mmol/L) = the numerical result
Glucose above 600 mg/dL (33,3 mmol/L) = HI**
*Repeat the test.
**Such results mean that the glucose concentration in the sample is very elevated and it indicates a serious risk. CONTACT A PHYSICIAN IMMEDIATELY.
Normal Values at fasting 70-110 mg/dL (3,9-6,1 mmol/L)
Packaging
Boxes of 25, 50 pcs


CHOLESTEROL STRIPS
Sample to be used
Fresh capillary whole blood
Range of Measurement
130-400 mg/dl (3,3-10,3 mmol/l)
Averange time for reading
About 30 seconds
Storage conditions
5-30 °C
Table of results
Cholesterol below 130 mg/dL (3.3 mmol/L) = LO*
Cholesterol between 130 mg/dL (3.3 mmol/L)
and 400 mg/dL (10.2 mmol/L) = the numerical result
Cholesterol above 400 mg/dL (10.2 mmol/L) the display = HI**
*Repeat the test.
**Such results mean that the Triglycerides concentration in the sample is very elevated and it indicates a serious risk. CONTACT A PHYSICIAN IMMEDIATELY.
Cholesterol Values
Less than 200 mg/dL (5.1 mmol/L) = desirable values
From 200 mg/dL (5.1 mmol/L) to 239 mg/dL (6.1 mmol/L) = high risk factor
More than or equal to 240 mg/dL (6.1 mmol/L) = high cholesterol
Packaging
Boxes of 5, 25 pcs


TRIGLYCERIDES STRIPS
Sample to be used
Fresh capillary whole blood
Range of Measurement
50-500 mg/dl (0,56-5,6 mmol/l)
Averange time for reading
About 30 seconds
Storage conditions
5-30 °C
Table of results
Below 50 mg/dL (0.5 mmol/L) = LO*
Between 50 mg/dL (0.5 mmol/L) and 500 mg/dL (5.6 mmol/L) = the numerical result
Above 500 mg/dL (5.6 mmol/L) = HI**
.*Repeat the test.
**Such results mean that the Triglycerides concentration in the sample is very elevated and it indicates a serious risk. CONTACT A PHYSICIAN IMMEDIATELY.
Triglycerides Values
Less than 150 mg/dl (1.7mmol/L) = Normal
Borderline High= 150-199 mg/dL (1.7-2.2 mmol/L)
High= 200-499 mg/dL (2.3-5.6 mmol/L)
Very High= 500 mg/dL (5.6 mmol/L) or above
Packaging
Boxes of 5, 25 pcs

HOW TO

MULTICARE IN
DIAGNOSTIC AID MULTIPARAMETER WITH CAPILLARY ASPIRATION

HOW TO PERFORM THE TEST OF GLUCOSE
1. Insert the data-chip
It is necessary to insert the data-chip contained in the strips box in order to automatically download all required information to perform the test.

2. Check the number
On the display will appear the data-chip number that should correspond to the one printed in the strip vial label.

3. Insert the strip
Insert the reagent strip of the parameter that you would perform. The instrument will automatically switch on.

4. Blood aspiration
After performing the finger prick let the blood be aspirated the drop of blood. The required volume is small.

5. Check the result
Check the result that appears on the display after short time. The wide characters will help the operator to execute the test and to clearly read the results.

6. Ejection of used strips
After performing the test it is possible to eject automatically the strip by pressing the switch on the backside of the instrument in order to avoid contamination.

HOW TO PERFORM THE TEST OF CHOLESTEROL AND TRIGLYCERIDES
1. Insert the data-chip
It is necessary to insert the data-chip contained in the strips box in order to automatically download all required information to perform the test.

2. Check the number
On the display will appear the data-chip number that should correspond to the one printed in the strip vial label.

3. Insert the strip
Insert the reagent strip of the parameter that you would perform. The instrument will automatically switch on.

4. Blood application
After performing the finger prick apply the drop of blood in the white circular area of the strip coming out the instrument. The quantity of blood should be enough to create a half-circle. Wait until moving segment will appear on the display

5. Check the result
Check the result that appears on the display after short time. The wide characters will help the operator to execute the test and to clearly read the results.

6. Ejection of used strips
After performing the test it is possible to eject automatically the strip by pressing the switch on the backside of the instrument in order to avoid contamination.
VETERINARY
BIOCHEMISTRY ANALYZER SEMI-AUTOMATIC
BIOCHEMISTRY ANALYZER AUTOMATIC