Laboratory photometer, completely Programmable up to 40 tests which can
be recalled by keyboard at any time. |
![]() |
Clin Check Plus is an interferential filter photometer,
designed for a precise and fast execution of the most important tests of
Clinical Chemistry and Hematology.
The Clin Check Plus
has 40 programs:
01 -26: programs are already set for the determination of
the analytes mentioned above;
27- 30: the last four are open which means
that new analyses can be introduced in them.
31- 40: Multistandard.
However, all the programs can be modified, except for
WBCH and WBTR, (Cholesterol and Triglyceride Tests on whole blood) due to
technical reasons.
All Clinical tests must be performed on serum except for
Cholesterol, Glucose and Triglycerides Tests, which can also be performed on
whole blood, using 10 µl of blood obtained, for example, by pricking a
finger.
Three Hematological parameters can be also be determined:
Erythrocytes, Hematocrit and Hemoglobin. Each test requires only 5
µl of whole blood.
Multistandard test up to 7 standard.
Standard value and concentrations are stored in non volatile memory.
The
execution of the analyses and programming the instrument are simple and guided
by the instructions shown on the display. It is possible to select a language
among German, Italian, French, English.
The light source of Clin
Check Plus is a 2 W tungsten lamp: this enables the instrument to read
in the UV range too. 5 interferential filters are incorporated: the proper
wavelength of the Test is automatically selected by microprocessors. Iit is
possible to add a sixth filter, upon request.
The instrument has a
2-position Dry Block Incubator at 37° C ; the Reading compartment is also at
37° C to ensure a constant temperature during the Reaction Time. The thermostat
turns on automatically when the instrument is switched on.
It is
possible to connect the Clin Check Plus to a serial printer or
to a computer for Data Elaboration.
All Clin Check Plus
instruments are subjected to complete Quality Control: electrical, photometric,
clinical and chemical controls by the manufacturer himself.
Comparing
various Clin Check Plus, you will note that the K-factor for
the Erythrocytes test is different between different instruments. The K-factor
for Hematocrit test varies also. The K-factors for the other tests remain the
same in all instruments. This is because Erythrocytes and Hematocrit are
determined turbidimetrically by means of the diffused beam at 90° ; the
intensity of such a beam can vary in a sensible way, depending on the optical
parts of each single instrument. Therefore, each single Clin Check
Plus is calibrated in our laboratories to guarantee a higher precision
in the results. The other analytes are determined by a colourimetric reaction
according to the Lambert-Beer Law.
The K-factors of Erythrocytes and
Hematocrit tests are indicated on a label at the back of each instrument.
Biochemical Systems International Srl reserves the right to change
specifications without prior notice.
| Spectral range | 320-750 nm |
| Filter Wavelenghts | 340-405-505-546-578-630 nm interferential filters |
| Operating modes | Absorbance, End Point, Fixed-Time, Kinetic and Multistandard |
| Measuring range | from -0.200 to +2.000 O.D. |
| Photometric accuracy | ± 1% from 0 to 2.000 O.D. |
| Photometric linearity | ± 1% from 0 to 2.000 O.D. |
| Reproducibility | CV < 1% @ 2.0 O.D. |
| Incubator | 2 positions at 37 °C; ± 0.2°C |
| Printer/Computer Output | RS-232 standard |
| Power Supply | AC or rechargeable batteries |
| AC Adapter | 220/110 V - 50/60 Hz |
| Dimension | 27.5 x 21 x 9 cm |
| Weight | 3.8 Kg |
Biochemical Systems International Srl reserves the right to change specifications without prior notice.
| CODE | REAGENT TYPE | METHOD | SAMPLE |
|---|---|---|---|
| HM3027 | Albumin | End-Point | Serum, Plasma |
| HM3012 | Alkaline Phosphatase | Kinetic | Serum, Plasma |
| HM3001 | Alpha Amylase | Kinetic | Serum, Plasma |
| HM3004 | Calcium | End-Point | Serum, Plasma |
| HM3006 | Chloride | End-Point | Serum, Plasma |
| HM3007 | Cholesterol | End-Point | Serum, Plasma |
| HM3022 | Cholesterol HDL | End-Point | Whole Blood |
| HM3032 | CK-MB | Kinetic | Serum, Plasma |
| HM3005 | Creatine Kinase | Kinetic | Serum, Plasma |
| HM3008 | Creatine | Fixed Time | Serum, Plasma |
| HM3011 | Erythrocytes* | Turbidimetric | Whole Blood |
| HM3017 | Gamma-GT | Kinetic | Serum, Plasma |
| HM3014 | Glucose | End-Point | Serum, Plasma |
| HM3015 | GOT | Kinetic | Serum, Plasma |
| HM3016 | GPT | Kinetic | Serum, Plasma |
| HM3009 | Hematocrit* | Turbidimetric | Whole Blood |
| HM3010 | Hemoglobin* | Colorimetric | Whole Blood |
| HM3013 | Inorganic Phosphorus | End-Point | Serum, Plasma |
| HM3023 | Iron | End-Point | Serum |
| HM3030 | Lactic Acid | End-Point | Whole Blood |
| HM3018 | LDH | Kinetic | Serum, Plasma |
| HM3035 | Magnesium | End-Point | Serum, Plasma |
| HM3019 | Proteins | End-Point | Serum, Plasma |
| HM3003 | Total Bilirubin | End-Point | Serum, Plasma |
| HM3020 | Triglycerides | End-Point | Serum, Plasma |
| HM3002 | Urea | End-Point | Serum, Plasma |
| HM3021 | Uric acid | End-Point | Serum, Plasma |
| HM3024 | WB Cholesterol* | End-Point | Whole Blood |
| HM3025 | WB Glucose* | End-Point | Whole Blood |
| HM3026 | WB Triglycerides* | End-Point | Whole Blood |
| HM3050 | Control Serum - Normal | ||
| HM3051 | Control Serum - Pathological |
Notes:
Liquid reagents prefilled in single cuvettes, 24 determinations per pack.
* Pack complete with capillanes
Biochemical Systems International Srl reserves the right to change specifications without prior notice.