Semi-automatic photometer for up to 120 different test programs
|
![]() |
3000 Evolution is an interferential filter photometer,
completely managed by microprocessors. It performs photometric measurements and
processes them according to programs with parameters that can be entered by the
operator. It executes, in a rapid and precise manner, most of important
chemistry and hematology tests.
Particularly, the following determinations
can be carried out:
- ABSORBANCE
- END-POINT
- KINETICS
-
FIXED-TIME
- MULTISTANDARD
- DIFFERENTIAL
Disposable macro and
micro cuvettes (glass or plastic), with an optical path of 1 cm, can be used by
simply removing the flow cell from the reading compartment and placing it on the
right-side one. Six filters are included in the instrument. The selection of the
interferential filter is automatic, with powered handling managed by a
microprocessor.
This feature makes reading easier and eliminates filter
selection errors that may occur in instruments which have a manual selection.
INSTRUMENT
DESCRIPTION
OPTICS
The optical part is very
sophisticated: it consists of a high-power halogen lamp (20 W) whose light beam
is centered by a quartz lens, thus allowing a high accuracy in measurements even
when reduced-volume cuvettes are used.
INCUBATOR
A
10-position dry incubator which can contain both square and cylindrical cuvettes
allow the sample incubation before reading. The temperature of the incubator is
equal to the one of the reading cell and can be regulated from 20 °C to 40 °C.
The execution of the analyses and instrument programming are simple and
performed by means of a keyboard, following the instructions shown on the
display. This display also shows the status and error/fault messages. The
analytical results are directly displayed in the measuring units selected by the
current program.
LANGUAGE The language of instructions
can be selected between English and Italian. Two other languages are avalaible
on request. The instrument is provided with a 24-column thermal printer that can
print analytical results as well as program parameters. All printed information
is sent to a serial RS-232 standard output. The printer can be disconnected and
data logged in PC. An advanced software guides and controls all operations
carried out by the user. An acoustic signal further helps the user, by emitting
a sound of a different tone from the usual one, it means you pushed the wrong
key.
MEMORY Up to 120 programs can be stored in the
EEPROM inside the instrument. The instrument is supplied already programmed.
Instrument is equipped with non volatile result memory. Up to 400 results can be
stored inside, for later or end-of-day recalling. Result may be recalled by
Test, Date or ID.Sample ID number is selectable and displayed before
testing.
The instrument is entirely
re-programmable
Biochemical Systems International Srl reserves
the right to change specifications without prior notice.
| Flow cell | 18 µL |
| Typical working volume | 500 µL |
| Minimum working volume | 350 µL |
| Carry over | less than 1% |
| Intake | peristaltic pump with programmable intake volume and air gap setting |
| Operating modes | Absorbance, End Point, Fixed-Time, Kinetic, Multistandard and Differential |
| Spectral range | 320-690 nm |
| Filter wavelength | 340-405-492-505-546-578-630 nm; 1 empty position |
| Reading | Monochromatic, Bichromatic |
| Reagent blank saving | Included |
| Measuring range | from -0.200 to 3.000 O.D. |
| Photometric accuracy | ± 1% from 0 to 2.500 O.D. |
| Photometric linearity | ± 1% from 0 to 2.500 O.D. |
| Reproducibility | CV < 1% from 0 to 2.000 O.D. |
| Results memory | 400 test results |
| ID sample | selectable |
| QC Program | last 30 results, 2 levels for 30 selectable tests with Levey-Jennings plot |
| Incubator | 10 positions; ± 0.2°C |
| Incubator temperature | selectable from 20°C to 40°C |
| Display | Graphic 240x128 pixels |
| Language | English, Italian, 2 other languages on request |
| Keyboard | 8 keys multifunctional keyboard or external connection for PS-2 keyboard |
| Printer | built-in graphic printer with 24 columns |
| Power supply | Autosensing (80-260 V) |
| External port | Serial RS232 |
| Dimension | 35x34x24 cm |
| Weight | 11 Kg |
Biochemical Systems International Srl reserves the right to change specifications without prior notice.
| CODE | REAGENT TYPE | METHOD | SAMPLE |
|---|---|---|---|
| HM3027 | Albumin | End-Point | Serum, Plasma |
| HM3012 | Alkaline Phosphatase | Kinetic | Serum, Plasma |
| HM3001 | Alpha Amylase | Kinetic | Serum, Plasma |
| HM3004 | Calcium | End-Point | Serum, Plasma |
| HM3006 | Chloride | End-Point | Serum, Plasma |
| HM3007 | Cholesterol | End-Point | Serum, Plasma |
| HM3022 | Cholesterol HDL | End-Point | Whole Blood |
| HM3032 | CK-MB | Kinetic | Serum, Plasma |
| HM3005 | Creatine Kinase | Kinetic | Serum, Plasma |
| HM3008 | Creatine | Fixed Time | Serum, Plasma |
| HM3011 | Erythrocytes* | Turbidimetric | Whole Blood |
| HM3017 | Gamma-GT | Kinetic | Serum, Plasma |
| HM3014 | Glucose | End-Point | Serum, Plasma |
| HM3015 | GOT | Kinetic | Serum, Plasma |
| HM3016 | GPT | Kinetic | Serum, Plasma |
| HM3009 | Hematocrit* | Turbidimetric | Whole Blood |
| HM3010 | Hemoglobin* | Colorimetric | Whole Blood |
| HM3013 | Inorganic Phosphorus | End-Point | Serum, Plasma |
| HM3023 | Iron | End-Point | Serum |
| HM3030 | Lactic Acid | End-Point | Whole Blood |
| HM3018 | LDH | Kinetic | Serum, Plasma |
| HM3035 | Magnesium | End-Point | Serum, Plasma |
| HM3019 | Proteins | End-Point | Serum, Plasma |
| HM3003 | Total Bilirubin | End-Point | Serum, Plasma |
| HM3020 | Triglycerides | End-Point | Serum, Plasma |
| HM3002 | Urea | End-Point | Serum, Plasma |
| HM3021 | Uric acid | End-Point | Serum, Plasma |
| HM3024 | WB Cholesterol* | End-Point | Whole Blood |
| HM3025 | WB Glucose* | End-Point | Whole Blood |
| HM3026 | WB Triglycerides* | End-Point | Whole Blood |
| HM3050 | Control Serum - Normal | ||
| HM3051 | Control Serum - Pathological |
Notes:
Liquid reagents prefilled in single cuvettes, 24 determinations per pack.
* Pack complete with capillanes
Biochemical Systems International Srl reserves the right to change specifications without prior notice.
| The user-friendly software and the graphic display make 3000 Evolution
very simple to use. The user can interact with the analyser by the 8-keys
keyboard on the instrument or by an external PS-2 keyboard. The User's manual explains how to execute tests and to customize execution modalities. The following picture shows the Main Window of the software: |
![]() |
| The following pictures show the display during test execution: |
![]() |
![]() |
| The first picture shows the display during the execution of a kinetic
analysis, the second one shows the result at the end of the
analysis. The software allows a lot of settings for the analyser, for example you can choose your favourite language. |
Biochemical Systems International Srl reserves the right to change specifications without prior notice.